    Monitor AMI CACE AC
    • AMI CACE is an economical, low-maintenance monitor that continuously measures conductivity - delivering reliability, efficiency and productivity for consistent measuring and gap free trend analyses. The on-line monitor AMI CACE continuously measures conductivity, before (specific / total conductivity) and after (acid / cation conductivity) cation exchange (CACE) as well as determining the pH value of the sample and alkalizing reagent based on differential conductivity measurement. With a measuring range of 0.055 to 1000 μS/cm, the AMI CACE is the perfect instrument dedicated to high quality conductivity measurements of feedwater, steam and condensate
      • Save money and maintenance down-time because no resin changing or resin rinse down time is require.​
      • Maximum instrument availability enables precise trend analyses giving you peace-of-mind.
      • Continuous monitoring of sample flow and sample temperature provides highest integrity of measured data
      • Straight forward instrument set-up allows rapid plant integration.
      • Experience of more than 25 years in water and steam quality measuring online instruments make AMI CACE a secure investment for you
    • Thông số kỹ thuật

      Transmitter AMI CACE in a rugged aluminum enclosure (IP 66).

      Swansensor UP-Con1000-SL; two 2-electrode conductivity sensors with slot-lock design and integrated Pt1000 temperature probe,k = 0.04 cm-1.

      Flow cell Catcon-Plus-SL CACE made of stainless steel 316L with digital sample flow meter. Quick sensor release with patented slot-lock design.

      EDI-Module with exchangeable sample chamber modul and automatic deaeration.

      Factory tested, ready for installation and operation.

      Dãy đo, độ chính xác tương ứng theo từng dãy đo như sau (tự động chuyển đổi range đo)
      Dãy đo: 0.055 ... 1'000 µS/cm
      Accuracy: ±1% of measured value or ±1 digit (whichever is greater)
      Power supply
      Voltage: 100 ... 240 VAC, 50/60Hz or 10-36 VDC
      Power consumption: max. 30 VA
      Sample conditions
      Sample conditions 3 to 4 L/h
      Temperature up to 50 °C (122 °F)
      Inlet pressure 0.5 bar
      Outlet pressure pressure free